Altitude Test Chamber of JOEO factory is designed to allow users to perform temperature, humidity, and altitude condition tests in a combined test chamber, and create multiple conditions to meet multiple test requirements at the same time.
Altitude chambers designed to simulate altitude conditions up to 100,000 feet.
Provides simultaneous temperature/altitude testing from site altitude to 65,000 feet for high altitude test chamber testing.
Humidity test can only be performed at the altitude of the site.
Heated inspection window to ensure that the test room is clearly visible.
Technical Indicators
- Temperature Range: -75°C to +150°C
- Pressure Range: Atmospheric pressure to 0.5kPa
- Pressure Deviation: ±0.1kPa
- Pressure Reduction Time: ≤40min (Atmospheric pressure to 0.1kPa)
- Temperature Fluctuation: ≤±0.5°C
- Temperature Uniformity: ≤±2.0°C
- Temperature Change Rate: ≥2.0 to 5.0°C/min
Main Features
- Flexible combination with various vacuum systems from domestic and international brands such as Leybold, Agilent, and Zhongke KeYi;
- Advanced pressure control system with small fluctuations and high stability;
- Single compressor refrigeration system with high efficiency, low energy consumption, and low operating costs;
- Compact structural design, lightweight, small footprint, convenient for transportation and installation;
- Noise reduction and vibration reduction technology, ultra-low noise and vibration;
- Performance indicators meet or exceed national standards, satisfying test procedures.
Standard:
Parameter
Model |
Temperature Low |
ACC-125 |
ACC-512 |
ACC-800 |
ACC-1000 |
ACC-2160 |
ACC-9000 |
Size |
Internal Dimension WxHxD (mm) |
500x500x500 |
800x800x800 |
1000x1000x800 |
1000x1000x1000 |
1800x1000x1200 |
1730x1980x2600 |
External Dimension WxHxD (mm) |
900x1680x1750 |
1200x1880x2050 |
1400x2080x2050 |
1400x2180x2250 |
2200x2180x2570 |
2130x2440x4673 |
|
Volume (L) |
125L |
512L |
800L |
1000L |
2160L |
8900L |
|
Pressure |
Pressure range |
Normal pressure~0.5Kpa ( 500pa) |
|||||
Pressure error |
Ordinary pressure~40kpa≦±2kpa; |
||||||
Depressurization time |
From site level to 15250m in 10minutes |
||||||
Rapid decompression time |
75.2kPa~18.8kPa≦15s (Optional) |
||||||
Explosive decompression time |
75.2kPa~18.8kPa≦0.1s (Optional) |
||||||
Temperature |
Temperature range |
Low temperature range: A:0℃; B: -20℃; C:-40℃; D:-70℃ |
|||||
Temperature fluctuation |
±0.5℃ |
||||||
Temperature Rising/ |
Temperature rising approx:0.1~ 2.0℃/min; |
||||||
Temperature uniformity |
±1.5℃(-40.0℃∽100.0℃) |
||||||
±2.0℃(100.1℃~180.0℃or-40.0℃∽-70.0℃) |
|||||||
Humidity |
Humidity range |
20%RH~98%RH(work under atmos pheric humidity) |
|||||
Humidity fluctuation |
±2.0%RH |
||||||
Humidity uniformity |
≦±3.0%RH(humidity >75%R.H) |
||||||
Power |
AC380V 3phase 5lines 50/60Hz |
||||||
Aprroximate weight(kg) |
570 |
1000 |
1200 |
1300 |
2500 |
7500 |
|
Marked |
Other dimensions accept customize, |